CCD Chip The highly sensitive 2/3” interline CCD image sensor used in the KeenviewG2 provides a resolution of 1376 x 1032 pixels at 14-bit dynamic range. The pixel size of the CCD chip is 6.45 x 6.45 µm² (effectively 23.5 x 23.5 µm²). Minimizing power consumption has made active camera cooling superfluous.
Binning Several binning modes are supported by KeenviewG2. The users can select between modes of 1x, 2x, 4x and 8x binning. This means sensitivity can be increased even more.
High frame rates The KeenViewG2 supports frame rates of more than 30 images per second at 2x binning and of 15 images per second at full resolution. This high frame rate is ideal for locating suitable sample segments directly onscreen.
CCD Cooling Noise is generally introduced if the CCD chip is not cooled to a sufficiently low temperature, known as “dark current”. This thermal noise is not a problem with the KeenViewG2 due to a sophisticated Peltier-cooling system which regulates the temperature of the CCD chip at 10° C (@ 25° C ambient temperature)
Variable exposure times The KeenViewG2 offers exposure times ranging from 100 µs to 160 seconds. This is a fundamental prerequisite for being able to acquire images of either intense illumination or of very low intensity.
Fiber optical coupling The KeenViewG2 camera system couples the electron-sensitive layer (scintillator) with the CCD chip using unique fiber optics (tapered). This innovative approach provides the user with a welcome price/performance advantage. In conjunction with the highly sensitive CCD chip and the optimized scintillator, this ensures high sensitivity and efficiency.
Optimized scintillators The KeenViewG2 uses a dedicated high-sensitivity phosphor scintillator. In order to match customers’ individual demands, various scintillators can be customized for several accelerating voltages and applications.
Anti-Blooming / Diffraction The CCD chip architecture guarantees extremely high anti-blooming. Individual pixels are clearly distinguished. In conjunction with the extremely short exposure times, this optimizes the acquisition of diffraction images
Housing Installing the KeenviewG2 is simple and straightforward. No additional water cooling and no other modifications at the TEM are necessary to set up the camera system. X-ray safety is guaranteed by the camera housing for up to 300 KeV.
Real-time functions Integration of the KeenViewG2 with iTEM, Olympus Soft Imaging Solutions TEM imaging platform ensures Real-Time shading correction, intensity histograms, auto contrast adjustment and Real-Time Fast Fourier Transformations during image acquisition.
Integration Furthermore, the integration of the camera with iTEM provides professional functions such as autofocus, labeling, processing, archiving, analyzing and report generation. Photographic quality printouts can be obtained virtually in seconds after acquisition – without the darkroom.
|